Test ID NSYPH Syphilis IgG Enzyme Immunoassay, Serum
Specimen Required
Collection Container/Tube: Serum gel
Acceptable: Red top
Submission Container/Tube: Plastic vial
Specimen Volume: 0.6 mL
Collection Instructions: Centrifuge and aliquot serum into plastic vial.
Useful For
An aid in the diagnosis of infection with Treponema pallidum
Routine prenatal screening
This test is not useful for diagnosis of congenital syphilis.
This test is not offered as a screening or confirmatory test for blood donor specimens.
Highlights
This testing should be used to assess for recent or past infection with Treponema pallidum or for routine prenatal screening.
Testing for syphilis is performed using the reverse screening algorithm at Mayo Clinic and Mayo Clinic Laboratories.
Method Name
Enzyme Immunoassay (EIA)
Reporting Name
Syphilis IgG EIA, SSpecimen Type
SerumSpecimen Minimum Volume
0.5 mL
Specimen Stability Information
Specimen Type | Temperature | Time | Special Container |
---|---|---|---|
Serum | Frozen | 14 days |
Reject Due To
Gross hemolysis | Reject |
Gross lipemia | Reject |
Heat-inactivated specimen | Reject |
Clinical Information
Syphilis is caused by infection with the spirochete Treponema pallidum subspecies pallidum. The infection is systemic, and the disease is characterized by periods of latency. These features, together with the fact that T pallidum cannot be isolated in culture, mean that serologic techniques play a major role in the diagnosis and follow-up of treatment for syphilis.
Historically, the serologic testing algorithm for syphilis included an initial non-treponemal screening test, such as the rapid plasma reagin (RPR) or the VDRL tests. Because these tests measure the host's immune response to nontreponemal antigens, they lack specificity. Therefore, a positive result by RPR or VDRL requires confirmation by a treponemal-specific test, such as the fluorescent treponemal antibody-absorption (FTA-ABS) or microhemagglutination (MHA-TP) assay. Although the FTA-ABS and MHA-TP assays are technically simple to perform, they are labor intensive and require subjective interpretation by testing personnel.
As an alternative to the traditional syphilis screening algorithm, many laboratories utilize the reverse syphilis screening algorithm. This algorithm starts with an automated treponemal assay to detect antibodies specific to T pallidum. If this screening assay is positive, the sample is reflexed for testing by RPR, which, if positive, is reported with a titer and is indicative of active or recent syphilis infection. If the RPR is negative, the sample is reflexed to a second treponemal assay, such as the T pallidum particle agglutination (TP-PA) assay. If the TP-PA is positive, this would indicate previously treated or late-stage syphilis infection. Alternatively, if the TP-PA is negative, the initial positive screen is interpreted as a false-positive result.
Syphilis screening at Mayo Clinic is performed using the reverse algorithm, which first tests sera for T pallidum specific IgG antibodies using an automated enzyme immunoassay. A positive treponemal test suggests infection with T pallidum but does not distinguish between recent, past, treated, or untreated infection. This is because treponemal tests may remain reactive for life, even following adequate therapy. Therefore, the results of a nontreponemal assay, such as RPR, are needed to provide information on a patient's disease state and history of therapy.(Table)
In some patients, the results of the treponemal screening test and RPR may be discordant (eg, syphilis IgG positive and RPR negative). To discriminate between a falsely reactive screening result and past syphilis, a second treponemal-specific antibody test is recommended using a method that is different from the initial screen test (eg, TP-PA).
In the setting of a positive syphilis IgG screening result and a negative RPR, a positive TP-PA result is consistent with either 1) past, successfully treated syphilis, 2) early syphilis with undetectable RPR, or 3) late/latent syphilis in patients who do not have a history of treatment for syphilis. Further historical evaluation is necessary to distinguish between these scenarios.(Table)
In the setting of a positive syphilis IgG screening result and a negative RPR, a negative TP-PA result is most consistent with a falsely reactive syphilis IgG screen.(Table) If syphilis remains clinically suspected, a second specimen should be submitted for testing.
Table. Interpretation and follow-up of reverse screening results:
Test and result |
|||||
Patient history |
Syphilis IgG Ab by EIA |
RPR |
TP-PA |
Interpretation |
Follow-up |
Unknown history of syphilis |
Nonreactive |
NA |
NA |
No serologic evidence of syphilis |
None, unless clinically indicated (eg, early/acute/primary syphilis) |
Unknown history of syphilis |
Reactive |
Reactive |
NA |
Untreated or recently treated syphilis |
See Centers for Disease Control and Prevention treatment guidelines |
Unknown history of syphilis |
Reactive |
Nonreactive |
Nonreactive |
Probable false-positive screening test |
No follow-up testing, unless clinically indicated (eg, acute/primary syphilis) |
Unknown history of syphilis |
Reactive |
Nonreactive |
Reactive |
Possible syphilis (eg, early or latent) or previously treated syphilis |
Historical and clinical evaluation required |
Unknown history of syphilis |
Equivocal |
NA |
NA |
NA |
Unknown history of syphilis |
Known history of syphilis |
Reactive |
Nonreactive |
Reactive or NA |
Past, successfully treated syphilis |
None |
EIA, enzyme immunoassay; NA, not applicable; RPR, rapid plasma reagin; TP-PA, Treponema pallidum particle agglutination |
Reference Values
Nonreactive
Reference values apply to all ages
Interpretation
Nonreactive:
No serologic evidence of exposure to Treponema pallidum (syphilis). Repeat testing may be considered in patients with suspected acute or primary syphilis.
Equivocal:
Recommend follow-up testing in 10 to 14 days if clinically indicated.
Reactive:
Results suggest infection with T pallidum at some point in time. Results do not distinguish between recent or past infection, or between treated and untreated syphilis as treponema-specific IgG may remain elevated despite appropriate therapy. Falsely reactive treponemal results may occur; additional testing by a non-treponemal assay is recommended if not previously performed on this sample.
Cautions
Despite active syphilis, serologic tests may be negative in severely immunosuppressed patients such as those with AIDS.
In very early cases of primary syphilis, serology tests may be negative.
In cases of untreated, late or latent syphilis, the result of rapid plasma reagin may be negative. However, the syphilis screening enzyme immunoassay (EIA) and Treponema pallidum particle agglutination (TP-PA) should be positive. A thorough clinical and historical evaluation should be performed to determine if treatment for latent syphilis is required.
Results should be considered in the context of all available clinical and laboratory data.
Clinical Reference
1. Centers for Disease Control and Prevention (CDC): Discordant results from reverse sequence syphilis screening-five laboratories, United States, 2006-2010. Morb Mortal WKLY Rep. 2011 Feb 11;60(5):133-137
2. Radolf JD, Tramont EC, Salazar JC: Syphilis (Treponema pallidum). In: Bennett JE, Dolin R, Blaser MJ, eds. Mandell, Douglas, and Bennett's Principles and Practice of Infectious Diseases. 9th ed. Elsevier; 2020:2865-289291
3. Binnicker MJ, Jespersen DJ, Rollins LO: Direct comparison of the traditional and reverse syphilis screening algorithms in a population with a low prevalence of syphilis. J Clin Microbiol. 2012 Jan;50(1):148-150
Day(s) Performed
Monday through Friday, Sunday
Report Available
Same day/1 to 3 daysTest Classification
This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.CPT Code Information
86780
NY State Approved
YesForms
If not ordering electronically, complete, print, and send Infectious Disease Serology Test Request (T916) with the specimen.